Code
Message
type/NA
64
Message/De-
scription
Cause
Measure
Priority
740 (A740)
Error C
factory set-
ting warn-
ing
C> Calculation
overflow, wrong
configuration
l
Mode level: The
measured pressure
has fallen short of the
value for HYDR.
PRESSURE MIN. or
exceeded the value
for HYDR. PRES-
SURE MAX.
l
Mode level: The
measured level had
decreased the value
for LEVEL MIN. or ex-
ceeded the value for
LEVEL MAX..
l
Mode flow: The
measured pressure
has exceeded the val-
ue for MAX. PRES-
SURE FLOW
l
Check parameter ad-
justment and carry out
a fresh adjustment, if
necessary
l
Select an instrument
with suitable measur-
ing range
l
Check parameter ad-
justment and carry out
a new adjustment, if
necessary
l
Check parameter ad-
justment and carry out
a fresh adjustment, if
necessary
l
Select an instrument
with suitable measur-
ing range
27
741 (A741)
Alarm B
B> TANK
HEIGHT outside
the editing limits
l
LEVEL MIN. or LEVEL
MAX. was changed
l
Carry out reset (Code
2710) and make a
fresh adjustment
44
742 (A742)
Alarm B
B> Initialisation
error of the sen-
sor
l
Electromagnetic influ-
ences are bigger than
the specifications in
chapter "Technical
data" (see product in-
formation manual VE-
GADIF 55). Normally
this message is only
displayed briefly. Ca-
ble connection sensor
- main electronics in-
terrupted.
l
Sensor defective
l
Wait several minutes
l
Carry out reset (Code
7864) and make a
fresh adjustment
l
Check cable connec-
tion and repair if nec-
essary
l
Exchange the sensor
18
743 (A743)
Alarm B
B> Error during
initialisation
l
Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data" (see product in-
formation manual VE-
GADIF 55), in general
this message is only
displayed briefly
l
Main electronics de-
fective
l
Wait several minutes
l
Restart the instrument.
Carry out a reset
(Code 62)
l
Exchange the main
electronics
14