In the following chart, all messages that can occur are listed.
Code
Message
type/NA 64
Message/De-
scription
Cause
Measure
Priority
101 (A101)
Alarm B
B> Checksum
error in the sen-
sor EEPROM
l
Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data" (see product in-
formation manual VE-
GADIF 55), in general
this message is only
displayed briefly
l
Sensor defective
l
Wait several minutes
l
Restart the instrument.
Carry out a reset
(Code 62)
l
Block electromagnetic
effects or remove the
source of interference
l
Exchange the sensor
17
102 (W102) Warning C
C> Checksum
error in the
pointer EE-
PROM
l
Main electronics de-
fective. As long as you
do not need the
pointer function,
measurement is cor-
rect and can be con-
tinued
l
Exchange the main
electronics
53
106 (W106) Warning C
C> Download
running - please
wait
l
Download running
l
Wait for the download
52
110 (A110)
Alarm B
B> Checksum
error in the con-
figuration EE-
PROM
l
Power supply is inter-
rupted during a write
operation
l
Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data" (see product in-
formation manual VE-
GADIF 55)
l
Main electronics de-
fective
l
Restore power supply,
if necessary, carry out
a reset (Code 7864)
and make a fresh ad-
justment
l
Block electromagnetic
effects or remove the
source of interference
l
Exchange the main
electronics
6
113 (A113)
Alarm B
B> ROM memo-
ry defective
l
Main electronics de-
fective
l
Exchange the main
electronics
1
115 (E115)
Error B fac-
tory setting
warning
B> Sensor over-
pressure
l
Existing overpressure
l
Sensor defective
l
Reduce pressure until
message disappears
l
Exchange the sensor
29
116 (W115) Alarm B
C> Download
faulty
l
The file is defective
l
The data are not
transferred correctly to
the processor during a
download, e.g. due to
open cable connec-
tions, voltage peaks
(ripple) in the voltage
supply or electromag-
netic effects
l
Use another file
l
Check cable connec-
tion PC - transmitter
l
Block electromagnetic
effects or remove the
source of interference
l
Carry out reset (Code
7864) and make a
fresh adjustment
l
Repeat download
36
120 (E20)
Error B fac-
tory setting
warning
C> Sensor low
pressure
l
Pressure too low
l
Sensor defective
l
Increase pressure until
message disappears
l
Exchange the sensor
30
Fault message via the
local display
Service instructions VEGADIF 55
7
Maintenance and fault rectification
31732
-
EN