beautypg.com

Fault message via the local display – VEGA VEGADIF 55 Service instructions User Manual

Page 7

background image

In the following chart, all messages that can occur are listed.

Code

Message
type/NA 64

Message/De-
scription

Cause

Measure

Priority

101 (A101)

Alarm B

B> Checksum
error in the sen-
sor EEPROM

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55), in general
this message is only
displayed briefly

l

Sensor defective

l

Wait several minutes

l

Restart the instrument.
Carry out a reset
(Code 62)

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange the sensor

17

102 (W102) Warning C

C> Checksum
error in the
pointer EE-
PROM

l

Main electronics de-
fective. As long as you
do not need the
pointer function,
measurement is cor-
rect and can be con-
tinued

l

Exchange the main
electronics

53

106 (W106) Warning C

C> Download
running - please
wait

l

Download running

l

Wait for the download

52

110 (A110)

Alarm B

B> Checksum
error in the con-
figuration EE-
PROM

l

Power supply is inter-
rupted during a write
operation

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55)

l

Main electronics de-
fective

l

Restore power supply,
if necessary, carry out
a reset (Code 7864)
and make a fresh ad-
justment

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange the main
electronics

6

113 (A113)

Alarm B

B> ROM memo-
ry defective

l

Main electronics de-
fective

l

Exchange the main
electronics

1

115 (E115)

Error B fac-
tory setting
warning

B> Sensor over-
pressure

l

Existing overpressure

l

Sensor defective

l

Reduce pressure until
message disappears

l

Exchange the sensor

29

116 (W115) Alarm B

C> Download
faulty

l

The file is defective

l

The data are not
transferred correctly to
the processor during a
download, e.g. due to
open cable connec-
tions, voltage peaks
(ripple) in the voltage
supply or electromag-
netic effects

l

Use another file

l

Check cable connec-
tion PC - transmitter

l

Block electromagnetic
effects or remove the
source of interference

l

Carry out reset (Code
7864) and make a
fresh adjustment

l

Repeat download

36

120 (E20)

Error B fac-
tory setting
warning

C> Sensor low
pressure

l

Pressure too low

l

Sensor defective

l

Increase pressure until
message disappears

l

Exchange the sensor

30

Fault message via the
local display

Service instructions VEGADIF 55

7

Maintenance and fault rectification

31732

-

EN

-080430