VEGA VEGADIF 55 Service instructions User Manual
Page 11
Code
Message
type/NA
64
Message/De-
scription
Cause
Measure
Priority
706 (W706) Warning C
C> Configura-
tions, data mem-
ory module and
instrument are
different
l
Configurations (pa-
rameter sets) in the
data memory module
and in the instrument
unequal
l
Copy data from device
into the data memory
module (see operating
instructions manual
VEGADIF 55, chapter
"C
opy configuration
data")
l
Copy data from device
into the memory mod-
ule (see also operating
instructions manual
VEGADIF 55, chapter
"C
opy configuration
data"). If the data
memory module and
the instrument have
different software ver-
sion, the message still
exists. The message
extinguishes when you
copy data from the
device into the data
memory module.
l
Instrument reset co-
des such as e.g. 7864
do not influence His-
toROM. I.e. if you
carry out a reset, the
configurations in His-
toROM and in the in-
strument can be un-
equal.
59
707 (W707) Alarm B
B> X-value of the
linearization
chart outside the
editing limits
l
At least one X-value of
the linearization chart
is either below the
value for HYDR.
PRESSURE MIN. or
MIN. LEVEL or above
the value for HYDR.
PRESSURE MAX. or
MAX. LEVEL.
l
Carry out a fresh ad-
justment (see operat-
ing instructions man-
ual VEGADIF 55)
38
710 (W710) Warning C
C> Set span is
smaller than per-
missible
l
Values for adjustment
(e.g. meas. start and
end) are too close
together
l
The sensor was ex-
changed and the cus-
tomer-specific param-
eter settings do not fit
the sensor
l
Unsuitable download
carried out
l
Adapt the adjustment
according to the sen-
sor (see operating in-
structions manual VE-
GADIF 55)
l
Replace the sensor
with another suitable
one
l
Check the parameter
adjustment and carry
out a fresh download
51
Service instructions VEGADIF 55
11
Maintenance and fault rectification
31732
-
EN
-080430