beautypg.com

5 instrument repair – VEGA VEGAFLEX 67 (-200…+400°C) 4 … 20 mA_HART two-wire User Manual

Page 43

background image

the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"S

ensor # online parameter adjustment". Connect this parameter

adjustment window before you carry out further steps.

S

elect with the right mouse key the sensor in the project and go to

"A

dditional function". Then click to "Software update". The window

"S

ensor # software update" opens. PACTware checks now the sensor

data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.

P

ush the button "Update software" and select the previously extracted

hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed
".

8

.5 Instrument repair

I

f a repair is necessary, please proceed as follows:

Y

ou can download a return form (23 KB) from our homepage at www.

vega.com under: "Downloads - Forms and certificates - Repair form".

B

y doing this you help us carry out the repair quickly and without

having to call back for needed information.

l

P

rint and fill out one form per instrument

l

C

lean the instrument and pack it damage-proof

l

A

ttach the completed form and, if need be, also a safety data

sheet outside on the packaging

l

P

lease ask the agency serving you for the address of your return

shipment. You can find the respective agency on our website
www.vega.com under: "Company - VEGA worldwide"

Load software into sen-
sor

VEGAFLEX

67 • 4 … 20

mA/HART - two-wire

43

8 M

aintenance and fault rectification

32311

-

EN

-120329