3 exchanging the electronics module – VEGA VEGAFLEX 67 (-200…+400°C) 4 … 20 mA_HART two-wire User Manual
Page 41
Error
Cause
Rectification
C
urrent signal
greater than 22 mA
or less than
3.6
mA
E
lectronics mo-
dule in the sensor
defective
E
xchange the instrument or send it in
for repair
I
n Ex applications, the regulations for the wiring of intrinsically safe
circuits must be observed.
T
he indicating and adjustment modules indicates faults via error codes
and text messages. The following table describes the error codes with
status according to NE 107 and gives information on the causes of
failure and their removal:
Status accor-
ding to NE 107
Error code
Text message
Cause/Rectification
F
ailure
E
013
no measured value available
S
ensor in boot phase
no measured value available
S
ensor does not find an echo, e.g.
due to faulty installation or wrong
parameter adjustment
no measured value available
W
rong sensor length entered
E
017
A
djustment span too small
A
djustment not within the specifica-
tion. Carry out the adjustment again,
increasing the distance between
min. and max. adjustment
E
036
N
o operable software
F
ailed or interrupted software up-
date/Repeat software update
E
042
H
ardware error, electronics defec-
tive
E
xchange the instrument or send it
in for repair
E
043
H
ardware error, electronics defec-
tive
E
xchange the instrument or send it
in for repair
8
.3 Exchanging the electronics module
I
f the electronics module is defective, it can be replaced by the user.
I
n Ex applications only one instrument and one electronics module with
respective Ex approval may be used.
I
f there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.
T
he new electronics module must be loaded with the settings of the
sensor. These are the options:
Error messages via the
indicating and adjust-
ment module
Sensor serial number
VEGAFLEX
67 • 4 … 20
mA/HART - two-wire
41
8 M
aintenance and fault rectification