beautypg.com

VEGA VEGASWING 61 - two-wire User Manual

Page 28

background image

28

9 Supplement

VEGASWING 61 • - two-wire

29227-EN-130409

Ʋ ECTFE

0.5 mm (0.02 in)

Ʋ PFA

0.5 mm (0.02 in)

Surface quality

Ʋ Standard

R

a

3 µm (1.18

-4

in)

Ʋ Hygienic version (3A)

R

a

< 0.8 µm (3.15

-5

in)

Ʋ Hygienic version (3A)

R

a

< 0.3 µm (1.18

-5

in)

Process fittings

Ʋ Pipe thread, cylindrical (DIN 3852-A) G¾ A, G1 A

Ʋ American pipe thread, conical

(ASME B1.20.1)

¾ NPT, 1 NPT

Ʋ Flanges

DIN from DN 25, ANSI from 1"

Ʋ hygienic fittings

Bolting DN 40 PN 40, Clamp 1" DIN 32676

ISO 2852/316L, Clamp 1½" PN 10 DIN 32676

ISO 2852/316L, conus DN 25 PN 40, Tuchenhagen

Varivent DN 50 PN 10

Max. torque - process fitting

Ʋ Thread G¾ A, ¾ NPT

75 Nm (55 lbf ft)

Ʋ Thread G1 A, 1 NPT

100 Nm (73 lbf ft)

High voltage test (enamel)

< 5 KV

Output variable
Output

Two-wire output

Output signal

Ʋ empty (uncovered)

8 mA

Ʋ full (covered)

16 mA

Ʋ Fault message

< 1.8 mA

Possible signal conditioning instruments VEGATOR 536Ex, 537Ex, 636Ex, VEGALOG 571
Modes (adjustable via the signal conditioning instrument)

Ʋ A

Max. detection or overflow/overfill protection

Ʋ B

Min. detection or dry run protection

Accuracy (according to DIN EN 60770-1)
Reference conditions and actuating variables according to DIN EN 61298-1

Ʋ Ambient temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Ʋ Product temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Product density

1 g/cm³ (0.036 lbs/in³) (water)

Ʋ Product viscosity

1 mPa s

Ʋ Superimposed pressure

0 kPa

Ʋ Sensor installation

Vertically from top

Ʋ Density selection switch

> 0.7 g/cm³