3 wiring plan, single chamber housing – VEGA VEGASWING 61 - two-wire User Manual
Page 15
15
5 Connecting to power supply
VEGASWING 61 • - two-wire
29227-EN-130409
8. Check the hold of the wires in the terminals by lightly pulling on
them
9. Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10. Screw the housing cover back on
The electrical connection is hence finished.
5.3 Wiring plan, single chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-d
version.
1
4
4
4
2
3
Fig. 8: Material versions, single chamber housing
1 Plastic (not with EEx d)
2 Aluminium
3 Stainless steel (not with EEx d)
4 Filter element for pressure compensation or blind plug with version IP 66/
IP 68, 1 bar (not with EEx d)
Z
+
0,5 g / cm3
0,7 g / cm3
11-36V DC
4
5
1
2
3
1
2
Fig. 9: Electronics and connection compartment
1 Control lamp
2 DIL switch for sensitivity adjustment
3 Ground terminal
4 Connection terminals
5 Processing system or PLC
We recommend connecting VEGASWING 61 in such a way that the
switching circuit is open when there is a level signal, line break or
failure (safe condition).
For connection to a VEGATOR signal conditioning instrument dto.
Ex, WHG. The sensor is powered by the connected VEGATOR signal
conditioning instrument. Further information is available in chapter
Housing overview
Electronics and connec-
tion compartment
Wiring plan