beautypg.com

VEGA VEGAPULS 51K…54K 4 … 20 mA; HART compact sensor User Manual

Page 55

background image

VEGAPULS 51K … 54K

55

21750-EN-031222

Technical data

Ambient conditions

Vessel pressure

VEGAPULS 51 and 52
- process fitting PVDF

-100 … 300 kPa (-1 … 3 bar)

- process fitting StSt

-100 … 1600 kPa (-1 … 16 bar)

VEGAPULS 53

-100 … 1600 kPa (-1 … 16 bar)

VEGAPULS 54
- PTFE coupling

-100 … 4000 kPa (-1 … 40 bar)

- ceramic coupling

-100 … 6400 kPa (-1 … 64 bar)

Ambient temperature on the housing

- 4 … 20 mA two-wire sensor

-40 … +80°C

- 4 … 20 mA four-wire sensor

-40 … +80°C

- 4 … 20 mA four-wire sensor Ex d ia

-40 … +60°C

Process temperature (flange temperature)

VEGAPULS 51
- process fitting PVDF

-20 … +80°C

- process fitting StSt

-40 … +120°C

VEGAPULS 52
- process fitting PVDF

-20 … +120°C (short-term 130°C)

- process fitting StSt

-40 … +150°C

VEGAPULS 53

-40 … +150°C

VEGAPULS 54
- PTFE coupling (standard)

-25 … +150°C

- PTFE coupling (option)

-15°C/-30°C/-40 … +150°C

- ceramic coupling

-60 … +250°C

Storage and transport temperature

-60 … +80°C

Protection

IP 66 and IP 67

Protection class

- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Ex-technical data

Comprehensive data in the safety instructions manual (yellow binder)

WHG approvals

All VEGAPULS 51 … 54 radar sensors are approved as part of an overfill protection sys-
tem for stationary vessels storing water-endangering liquids.