Ohaus T31XW Indicator Manual en User Manual

Page 21

background image

EN-19

3000 Series Indicators

Reset No, Yes
Legal for Trade

Off, On

Cal Unit kg, lb
Capacity

5…20000

Graduation

0.001…20

Power On Unit g, kg, lb, oz, lb:oz, Auto
Zero Range 0%, 2%, 100%
End Setup Exit SETUP menu

3.4 Setup Menu

When the Indicator is used for the first time, enter this
menu to set the Capacity and Graduation.

3.4.1 Reset

Reset the Setup menu to the factory defaults.
No

= not reset.

Yes = reset.

NOTE: If the Legal for Trade menu item is set to ON, the Capacity, Graduation, Zero Range and Legal For
Trade settings are not reset.

3.4.4 Capacity

Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.

3.4.2 Legal for Trade

Set the legal for trade status.
OFF = off
ON = on
Turning on the “LFT” menu setting has the following effects:
• Zero-range is set and locked on “2”.
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.

3.4.3 Calibration Unit

Set the unit during calibration.
CAL UN kg = Calibrate using kg weights
CAL UN lb = Calibrate using pound weights