Calibration and sealing, 1 calibration – Ohaus RE PRICE COMPUTING COMPACT SCALES Manual User Manual

Page 15

background image

RE Series

EN-13

6. CALIBRATION AND SEALING


6.1
Calibration

For best results, calibrate the scale at regular intervals. Temperature
changes, geographic gravity variations, altitude changes and abuse are few
reasons why a scale may need recalibration.


Note: The following procedure requires the use of hand tools and calibration
weights. If unable to perform the calibration, contact an authorized Ohaus
service organization.


1. With the scale on, remove any sticker or sealing screw (1), then gently

press the switch inside the access hole. A narrow screwdriver may be

inserted through the hole to press the switch.

Note: When accessing the bottom of the scale, avoid placing the

scale upside down on the pan or sub-platform. Place the scale on its
side.

2. “

S1 oFF

is displayed momentarily. Press

[PLU] [2] [4] [6] [8] [1]

[3].


3. “

SEt

” is displayed. Press

[FIX] successively to move through the

succeeding parameters: “

GEo”,

SingLE”,

CAL n”.

Note: Press [>0<] at any time to return to the previous parameter, or

the scale may be turned off at any time to abort setup without saving
any changes.


4. When “

CAL n

” is displayed, press

[>T<] to change the setting to

CAL Y

”.


5. When “

CAL Y

” is displayed, press

[FIX] to enter the Span Calibration

sequence.