background image

EN-2

CHECKWEIGH BASE

2. INSTALLATION

2.1 Unpacking

Unpack and inspect the product to make sure that all components have been included. The CKW Base package
includes the following:

CKW Base

Weighing Pan

Warranty Card

Instruction Manual

When purchased as a complete CKW scale, will also include:

CKW-55 Indicator

Column Assembly

Check to ensure the product was not damaged during shipping. Immediately inform your OHAUS dealer in the
unlikely event that you find damaged or missing parts.

Save the packaging material. Re-pack the product using original packaging materials to ensure safe storage or
transport.

2.2 Assembly

2.2.1 Weighing Pan

Place the weighing pan securely over the rubber load pads on the top plate of the base (first ensure that all load
pads are seated fully in their respective mounting holes).

2.2.2 Wiring Connections

Connect the load cell cable to an indicator using the wiring codes in Tables 2-1 and 2-2 below. Refer to the
indicator manual for information about load cell connections, setup and calibration.
Note: When purchased as a CKW scale, the base is already pre-wired to the CKW-55 Indicator.

TABLE 2-1. FOUR WIRE LOAD CELL CONNECTION

(CKW3R, 6R and 15L)

TABLE 2-2. SIX WIRE LOAD CELL CONNECTION

(CKW30L only)

Function

Wire Color

Function Wire Color

+ Excitation

Green

+ Excitation Green

- Excitation

Black

- Excitation

Black

+ Signal

Red

+ Signal

Red

- Signal

White

- Signal

White

+ Sense

Blue

- Sense

Brown

Shield

Yellow