beautypg.com

14 ramping functions example, In section 3.14 on, Operation – American Magnetics 4Q06250PS-430 High Stability Integrated Power Supply System User Manual

Page 103

background image

Rev. 5

85

Operation

Ramping Functions Example

2. Press the

SHIFT

and then the

PERSIST. SWITCH

CONTROL

key to turn on the persistent switch heater

current. Note that the Model 430 Programmer will enter the

HEATING SWITCH mode and disallow any ramping during

the switch heating period.

3. At the end of the switch heating period, the

MAGNET IN

PERSISTENT MODE

LED will be turned off and the Model

430 Programmer will be in the PAUSED mode.

3.14 Ramping Functions Example

As an example of ramping to two target field/current settings, refer to

Figure 3-14 below. Each step is labeled as 1 through 8 in Figure 3-14. The

Model 430 Programmer, for the purposes of the example, is assumed to be

in the PAUSED mode at 0 A at the beginning of the ramping example.

Point 1. The current is 0 A and the Model 430 Programmer is in the

PAUSED mode. The user sets the target field/current to -40.000 A. The

RAMP / PAUSE

key is pressed so that the PAUSED mode is no longer

active and the Model 430 begins ramping current.

Point 2. The target field/current setting of -40.000 A is achieved and the

Model 430 Programmer switches to HOLDING mode.

Point 3. The user changes the ramp rate setting. The user also sets a new

value of +40.000 A for the target field/current. As soon as the new target

Figure 3-14.

Ramping to Two Different Target Field/Current Settings.