Installation
21
Checking Voltage and Connecting Power
Check that the voltage available at the power source is within specification before
connecting it to the readout. If required, set the voltage selector to match the line
voltage.
1.
Remove the caution label from the input
module, and use a thin-blade screwdriver in
the slot at the top of the power-input module
to open the module cover.
2.
Pull the selection block out.
3.
Remove the fuse, move the shorting bar to the other
side, install the proper fuse, and push the block
back into place.
4.
Close and snap the cover shut.
The voltage setting will show through the window in the voltage selection cover.
Connect the readout to the power source using the power cord supplied.
Opening
slot
!
CAUTION
Connecting the readout to a power source outside of the
acceptable range, or making an inappropriate setting with the
voltage selector, may damage the readout, the encoders, or
both.