• Check function of the linear encoder.

These steps will confirm proper operation of your installation.
Counting tests check channel A and B output function and Reference
Mark tests check Reference Mark operation.

• Configure the readout for proper counting resolution.
• Move axis and compare the display to the movement.
• Repeat test for mechanical integrity and repeatability.
• Configure the readout for sensing reference mark.
• Move the reading head approximately 20mm.
• Locate an indicator on scale end and zero out axis and indicator.
• Move axis out and back over full travel and return to dial zero.
• Electronics should read zero ± 1 count.

ACU-RITE

®

10

Checking the Installation

MicroScale

• Zero the display and indicator.
• Move axis to the end of it’s travel, then return to dial zero.
• Readout should read zero ± 1 count.

Repeatability Test

Place indicator at end of

moving component (scale case

shown, or reading head).

• Route the cables with slack loops to allow for axis motion.
• Secure excess cable by fastening with clips or ties.
• Attach the linear encoder connectors to the readout.

Readout

Counting Test

Connecting

Secure excess

cable

Provide slack

loops

Attach connector