ACTION
Change batteries.
See Page 48 immediately.
If this display appears
before blood is placed on
strip, remove test strip and reinsert. If this display
appears
after blood is applied to the strip, there is
an incorrect amount of blood on the strip. Repeat
the test with a new test strip.
Remove Chek Strip, turn meter OFF and repeat qual-
ity control check. See page 52,
Quality Control
Testing. If this error display still appears, testing
cannot take place. Call Accu-Chek Customer Care
SM
at 1-800-858-8072.
Other Information
95
94
DISPLAY
CAUSE
Battery is dead. Only
memory results can be
recalled.
Blood glucose result
may be extremely high
and above the system’s
reading range.
Test strip may be dam-
aged or test was not
performed correctly.
Meter has failed its sys-
tem check.