If the result is within the acceptable range:
• You can feel confident that your test strips and
meter are working correctly.
If the result is outside the acceptable range:
• Check the expiration date of the test strips and
control solutions. If the strips or control solution
are expired, dispose of them promptly.
• Make sure the cap was on the vial of test strips.
• Check if the code in the meter matched the code
on the test strip vial label.
• Make sure you followed the testing steps exactly.
• Repeat the glucose control test with a new test
strip.
If the result is still not within the acceptable range,
your blood glucose results may not be reliable.
Contact Accu-Chek Customer Care
SM
for help at
1-800-858-8072, 24 hours a day.
65
Understanding Your Quality Control Results
The test strip vial label indicates the acceptable
range for Low Level and High Level glucose control
solutions. The result displayed on your meter should
fall within the acceptable range for the level of solu-
tion you used.
64
Quality Control T
esting
362-11463 AdvMan Classic-ENGL 7/15/02 12:05 PM Page 64