Getting Started

28

AD007 User Manual

To connect the AD007 GPIB-LAN Adapter to the GPIB, do these
steps:

1.

Connect the GPIB port of the adapter to the GPIB port(s) or
connector(s) for the instrument(s) that you wish to control. If
necessary, use the GPIB connector hardware provided. (See
Figure 8.)

129-1502-00

GPIB cable

GPIB port on
instrument

Figure 8: Connecting to the GPIB

2.

Turn on all devices connected to the GPIB. Wait for all power on
routines to finish.

3.

Access the web page on the AD007 adapter.

4.

Click GPIB Communications.

5.

Click Poll to refresh the list of active talk/listen instruments.

6.

Verify that the devices on the bus are listed.

7.

If you change which devices are active on the bus or device
parameters, click Poll to rediscover the active addresses on the
GPIB.