beautypg.com

ResMed VPAP III & III ST User Manual

Page 48

background image

42

Recommended separation distances between portable and mobile RF communications
equipment and the VPAP

The VPAP is intended for use in an environment in which radiated RF disturbances are controlled. The

customer or the user of the VPAP can help prevent electromagnetic interference by maintaining a minimum

distance between portable and mobile RF communications equipment (transmitters) and the VPAP as

recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter
m

Rated maximum output

power of transmitter

W

150 kHz to 80 MHz

d = 1.17

P

80 MHz to 800 MHz

d = 0.35

P

800 MHz to 2.5 GHz

d = 0.35

P

0.01

0.17

0.04

0.04

0.1

0.37

0.11

0.11

1

1.17

0.35

0.35

10

3.69

1.11

1.11

100

11.70

3.50

3.50

For transmitters rated at a maximum output power not listed above, the recommended separation distance

d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where

P is the maximum output power rating of the transmitter in watts (W) according to the transmitter

manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.