Pressure variation – ResMed VPAP III & III ST User Manual
Page 42
![background image](/manuals/197087/42/background.png)
36
IEC 60601-1 Classifications
Class II (double insulation)
Type CF
Table 7: Displayed values
* Results may be inaccurate in the presence of leaks.
** The displayed values are estimates. They are provided for trending purposes only.
Pressure Variation
Value
Range
Accuracy
Display Resolution
Pressure sensor at air outlet
Pressure
-5 to 30 cm H
2
O
±0.5 cm H
2
O
0.1 cm H
2
O
Flow sensor in flow generator*
Leak
0–120 L/min
**
1 L/min
Tidal volume
100–3000 mL
**
1 mL
Respiratory rate
6–60 BPM
±0.5 BPM
0.1 BPM
Minute ventilation
0.6–60 L/min
**
0.25 L/min