English – ResMed VPAP III & III ST User Manual

Page 47

background image

41

System Specifications

English

Guidance and manufacturer’s declaration – electromagnetic immunity (Continued)

The VPAP is intended for use in the electromagnetic environment specified below. The customer or the

user of the VPAP should assure that it is used in such an environment.

Immunity
test

IEC60601-1-2
test level

Compliance
level

Electromagnetic environment –
guidance

Portable and mobile RF communications

equipment should be used no closer to any

part of the VPAP, including cables, than the

recommended separation distance

calculated from the equation applicable to

the frequency of the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

3 Vrms

d = 1.17

P

Radiated RF

IEC 61000-4-3

10 V/m

80 MHz to 2.5 GHz

10 V/m

d = 0.35

P 80 MHz to 800 MHz

d = 0.70

P 800 MHz to 2.5 GHz

where P is the maximum output power

rating of the transmitter in watts (W)

according to the transmitter manufacturer

and d is the recommended separation

distance in meters (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site
survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following

symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an

electromagnetic site survey should be considered. If the measured field strength in the location in which

the VPAP is used exceeds the applicable RF compliance level above, the VPAP should be observed to verify

normal operation. If abnormal performance is observed, additional measures may be necessary, such as

reorienting or relocating the VPAP.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.