beautypg.com

ResMed VPAP III & III ST User Manual

Page 31

background image

25

How to Use the VPAP Menus

English

SmartStart™

The VPAP has a function called SmartStart which can be enabled by your
clinician. If SmartStart is enabled, VPAP will start automatically when you breathe
into the mask and will stop automatically when you take your mask off. This
means you do not have to press the Front key to begin or end treatment.
Note: If you select “Mir Full” as the mask option, SmartStop is automatically
disabled. SmartStart may not work with an Ultra Mirage Full Face Mask due to
safety features of the mask.
When Mask Alarm is set to ON, SmartStart/Stop automatically reverts to OFF.
SmartStop cannot be used with Mask Alarm because if a high leak occurs,
SmartStop will stop treatment before the Mask Alarm signal is activated.

Settings for Mask Types

The following table shows the setting that should be selected for each mask
type.

Table 3: Settings for mask types

SmartStart

OFF

Turns the SmartStart function on
or off.
See “SmartStart™” on page 25.

ON/OFF

Mask Alarm

ON
(VPAP III ST);
OFF
(VPAP III)

Turns the Mask Alarm signal on or
off.
If enabled, the Mask Alarm feature
will alert you when a high mask
leak is detected. An audible tone
will sound and a high leak
message will appear on the LCD
screen.

ON/OFF

Function

Default

Function Description

Setting Options

Settings

Mask

MIRAGE

Mirage Nasal Mask

ULTRA

Ultra Mirage Nasal Mask

STANDARD

Mirage Activa Nasal Mask
Mirage Vista Nasal Mask
Modular Nasal Mask

MIR FULL

Mirage Full Face Mask
Mirage Full Face Mask Series 2
Ultra Mirage Full Face Mask