Power Requirements

Current consumption from USB.............500 mA, max

Suspend mode..................................2.5 mA, max

Bus Interface

USB specification ...................................USB 2.0 Hi-Speed

Physical Characteristics

Dimensions .............................................12.1 cm

× 8.6 cm × 2.5 cm

(4.75 in.

× 3.37 in. × 0.99 in.)

Weight ....................................................Approx. 250 g (8.8 oz)

Safety

If you need to clean the module, wipe it with a dry towel.

Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-earth ground.........................±30 V max,

Measurement Category I

Isolation

Channel-to-channel..........................No isolation between channels

Channel-to-earth ground..................No isolation between channels

and earth ground

Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.

Caution

Do not connect the NI USB-9234 to signals or use for measurements within

Measurement Categories II, III, or IV.