beautypg.com

National Instruments NI 9233 User Manual

Page 24

background image

NI 9233 Operating Instructions and Specifications

24

ni.com

Total harmonic distortion (THD)

Intermodulation distortion (–1 dBFS)

DIN 250 Hz/8 kHz
4:1 amplitude ratio ..................... –80 dB
CCIF 11 kHz/12 kHz
1:1 amplitude ratio ..................... –93 dB

MTBF ............................................... 397,465 hours at 25 °C;

Bellcore Issue 2, Method 1,
Case 3, Limited Part Stress
Method

Note

Contact NI for Bellcore MTBF specifications

at other temperatures or for MIL-HDBK-217F
specifications.

Input

Amplitude

1 kHz,

–40 to 70 °C

10 kHz,

25 to 70 °C

10 kHz,

–40 to 25 °C

–1 dBFS

–90 dB

–80 dB

–80 dB

–20 dBFS

–95 dB

–90 dB

–80 dB