beautypg.com

National Instruments NI 9233 User Manual

Page 20

background image

NI 9233 Operating Instructions and Specifications

20

ni.com

Common-mode voltage range
(AI– to earth ground) ........................ ±2 V max

IEPE excitation current

Minimum.................................... 2.0 mA
Typical ........................................ 2.2 mA

IEPE compliance voltage.................. 19 V max

Use the following equation to make sure that your configuration
meets the IEPE compliance voltage range.

V

common-mode

+ V

bias

±V

full-scale

must be 0 to 19.

where

V

common-mode

is any common-mode voltage applied to

the NI 9233,

V

bias

is the bias voltage of the accelerometer, and

V

full-scale

is the full-scale voltage of the accelerometer.