beautypg.com

National Instruments 24-Bit Half/Full-Bridge Analog Input Module NI 9237 User Manual

Page 23

background image

© National Instruments Corp.

23

NI 9237 Operating Instructions and Specifications

Overvoltage protection
between any two pins........................ ±30 V

Accuracy

Gain drift........................................... 10 ppm/°C max

Offset drift

2.5 V excitation .......................... 0.6

μV/V per °C

3.3 V excitation .......................... 0.5

μV/V per °C

5 V excitation ............................. 0.3

μV/V per °C

10 V excitation ........................... 0.2

μV/V per °C

Measurement Conditions

*

Percent of

Reading

(Gain Error)

Percent of

Range

**

(Offset Error)

Calibrated typ (25 °C, ±5 °C)

0.05%

0.05%

Calibrated max (–40 to 70 °C)

0.20%

0.25%

Uncalibrated typ (25 °C, ±5 °C)

0.20%

0.1%

Uncalibrated max (– 40 to 70 °C)

0.60%

0.35%

*

Before offset null or shunt calibration.

**

Range equals 25 mV/V.