beautypg.com

Calibration, Bus interface, Power requirement – National Instruments NI PXI-7831R User Manual

Page 51

background image

Appendix A

Specifications

NI PXI-7831R User Manual

A-6

ni.com

Timebase accuracy

With onboard base clock .................±100 ppm

Phase locked
to PXI 10 MHz clock.......................±350 ps jitter, 300 ps skew (max)

Calibration

Recommended warm-up time.................15 minutes

Calibration interval .................................1 year

Onboard calibration reference

DC level...........................................5.000 V (±3.5 mV)

(actual value stored
in flash memory)

Temperature coefficient...................±5 ppm/°C max

Long-term stability ..........................±20 ppm/

Note

To generate a calibration certificate for the NI PXI-7831R, click On-line

Calibration Certificates at

ni.com/calibration

.

Bus Interface

PXI..........................................................Master, slave

Power Requirement

+5 VDC (±5%)

NI PXI-7831R .................................450 mA (typ), 700 mA (max)

(does not include current drawn
from the +5 V line on the
I/O connectors)

+3.3 VDC (±5%)

NI PXI-7831R .................................335 mA (typ), 730 mA (max)

Power available at I/O connectors ..........+4.65 to +5.25 VDC at 1 A total,

250 mA per I/O connector pin

1 000 h

,