Maximum working voltage, In appendix, Accuracy information dc transfer characteristics – National Instruments NI 783xR User Manual
Page 47: Dynamic characteristics
![background image](/manuals/191151/47/background.png)
Appendix A
Specifications
A-2
ni.com
Maximum working voltage
(signal + common mode) ........................Inputs should remain
within
±12 V of ground
Overvoltage protection
...........................±42 V
Data transfers ..........................................Interrupts, programmed I/O
Accuracy Information
DC Transfer Characteristics
INL
..........................................................±3 LSB typ, ±6
LSB max
DNL ........................................................–1.0 to +2.0 LSB max
No missing codes resolution...................16 bits typ, 15 bits min
CMRR, DC to 60 Hz ..............................86 dB
Dynamic Characteristics
Bandwidth
Small signal (–3 dB)........................650 kHz
Large signal (1% THD) ...................55 kHz
System noise ...........................................1.8 LSB
rms
(including quantization)
Nominal Range (V)
Absolute Accuracy
Relative
Accuracy
% of Reading
Offset
(
µV)
Noise +
Quantization
(
µV)
Temp
Drift
(%/
°C)
Absolute
Accuracy
at Full
Scale
(
±
mV)
Resolution (
µV)
Positive
Full
Scale
Negative
Full
Scale
24
Hours
1 Year
Single
Point
Averaged
Single
Point
Averaged
10.0
–10.0
0.0496
0.0507
2542
1779
165
0.0005
7.78
2170
217
Note: Accuracies are valid for measurements following an internal calibration. Measurement accuracies are listed for
operational temperatures within
±
1 °C of internal calibration temperature and
±
10 °C of external or factory-calibration
temperature. Temp drift applies only if ambient is greater than
±
10 °C of previous external calibration.