beautypg.com

Electromagnetic conformity – Philips 861304 User Manual

Page 58

background image

50

Ph
ilip

s M
e

di

ca

l Sy
st

e

m

s

HEARTSTART FRx 861304 OWNER’S MANUAL

AP

P

E

ND

IC

ES

H

Additional technical data required for European conformity

Electromagnetic conformity

Guidance and manufacturer’s declaration: The HeartStart FRx is intended for use
in the electromagnetic environment specified in the tables below. The customer or
user of the HeartStart FRx should assure that it is used in such an environment.

Electromagnetic emissions

emissions test

compliance

electromagnetic environment – guidance

RF

CISPR 11

Group 1

Class B

The FRx uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The FRx is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.