Philips 861304 User Manual
Page 42

34
Ph
ilip
s M
e
di
ca
l Sy
st
e
m
s
HEARTSTART FRx 861304 OWNER’S MANUAL
AP
P
E
ND
IC
ES
Controls and indicators
sealing
Meets IEC 529 class IPx5 for jetting water and class IP5x for solid objects (dust
protected).
ESD/EMI (radiated and
immunity)
See Appendix F.
aircraft: method
Meets RTCA/DO-160D:1997 Section 21 (Category M - Charging).
category
specifications
category
specifications
controls
Green On/Off button
Blue i-button
Orange Shock button
Optional Infant/Child Key accessory
indicators
Ready light: green, blinks when the defibrillator is in standby mode (ready for
use); solid when the defibrillator is being used.
i-button: blue, flashes when information is available, on solid during patient care
pause.
Caution light: flashes when the defibrillator is analyzing, comes on solid when
the defibrillator is ready to deliver a shock.
Shock button: orange, flashes when the defibrillator is charged and ready to
deliver a shock.
Pads Placement LEDs: flash when FRx is turned on; off once pads are placed on
patient. Also operates with Infant/Child Key inserted to indicate pads placement
on infants and children under 55 pounds (25 kg) or 8 years old.
audio speaker
Provides voice instructions and warning tones during normal use.
beeper
Provides chirps when troubleshooting is needed.
status indicator
Status indicator LCD displays device readiness for use.
low battery detection
Automatic during daily periodic self-testing.
low battery indicator
Alarm chirps and flashing blue in-button.