Technical information
HeartStart FRx 861304 Defibrillator specifications
The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manual for HeartStart
Defibrillators, located online at www.medical.philips.com.
Physical
Environmental
category
specifications
size
2.4” H x 7.1” D x 8.7” W
(6 cm H x 18 cm D x 22 cm W).
weight
Approximately 3.5 lbs (1.6 kg) with battery and pads installed.
pads compatibility
HeartStart SMART Pads II 989803139261
(In an emergency or during use, HeartStart DP2/DP6 pads may be used.
However, the FRx should not be stored with DP2/DP6 pads installed, as the
daily self-test will not give a “pass” result and the device will chirp.)
category
specifications
temperature and
relative humidity
Operating and standby (battery installed, pads connected):
32° to 122° F (0° to 50° C);
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads case):
-4° to 140° F (-20° to 60° C) for up to 1 week;
0% to 85% RH (non-condensing) for up to 2 days, thereafter 65% RH maximum
altitude
0 to 15,000 feet (0 to 4,572 m).
shock/drop abuse
tolerance
Withstands 1 meter drop on any edge, corner, or face of the device onto
masonry surface.
vibration
Operating: meets MILSTD 810F Fig. 514.5C-17, random.
Standby: meets MILSTD 810F Fig. 514.5C-18, swept sine (helicopter).