HEARTSTART FRx 861304 OWNER’S MANUAL

AP

P

E

ND

IC

ES

energy

Using HeartStart SMART Pads II for adult defibrillation: 150 J nominal (±15%)
into a 50 ohm load.

Using HeartStart SMART Pads II with Infant/Child Key inserted: 50 J nominal
(±15%) into a 50 ohm load. Sample pediatric energy doses:

age

energy dose

newborn

14 J/kg

1 year

5 J/kg

2 - 3 years

4 J/kg

4 - 5 years

3 J/kg

6 - 8 years

2 J/kg

Doses indicated are based on CDC growth charts for the 50th percentile
weights for boys.*

* National Center for Health Statistics in collaboration with the National Center for

Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age
percentiles, revised and corrected
November 28, 2000. Atlanta, GA: Centers for Disease
Control and Prevention © 2000.

charge control

Controlled by Patient Analysis System for automated operation.

shock-to-shock cycle time

< 20 seconds typical, including analysis.

“charge complete”

indicator

Shock button flashes, audio tone sounds; device is able to deliver a shock as
soon as a shock is advised.

patient care

pause-to-shock time

Quick Shock. 8 seconds, typical, from end of patient care pause to shock
delivery.

disarm (AED mode)

Once charged, the HeartStart FRx will disarm if:
• patient’s heart rhythm changes to non-shockable rhythm,
• a shock is not delivered within 30 seconds after the FRx is armed,
• the On/Off button is pressed for one second to turn off the FRx,
• the Infant/Child Key is inserted or removed,

the battery has been removed or is completely depleted, or

• the impedance between pads is out of range.

adult shock delivery vector

Via SMART Pads II placed in the anterior-anterior (Lead II) position.

infant/child shock delivery

vector

Via SMART Pads II typically placed in the anterior-posterior position.