If the FRx does not turn on when you press the On/Off button, remove the
battery and replace it with a new battery if available and press the On/Off button
to turn on the defibrillator. If no spare battery is available, remove the installed
battery for five seconds, then reinsert it and run a battery insertion self-test.

If the problem continues, do not use the defibrillator. Attend to the patient,
providing CPR if needed, until emergency medical personnel arrive.

Troubleshooting while the FRx is being used

(green Ready light is solid)

Always follow any instructions the device gives.

defibrillator says:

possible cause

recommended action

... to replace the

battery immediately

The battery is nearly depleted. The
FRx will turn off unless a new battery
is installed.

Install a new battery immediately.

... to plug in pads

connector

... to replace pads

• The pads connector has been

unplugged.

• The pads have been damaged.
• The pads have been peeled from the

case, but have not been successfully
attached to the patient. There may
be a problem with the pads.

• Plug in the pads connector.

• Replace the damaged pads.
• Replace the pads on patient with

new pads to continue with the
rescue.

... to press the pads firmly

to the skin

... to make sure the pads

have been removed

from the case

... the pads should not

be touching the

patient’s clothing

... to make sure the pads

connector is fully inserted

• The pads are not properly applied to

the patient.

• The pads are not making good

contact with the patient's bare chest
because of moisture or excessive
hair.

• The pads are touching each other.
• The pads may not have been

removed from the case or may be on
the patient’s clothing.

• Pads connector is not fully inserted.

• Make sure that the pads are sticking

completely to the patient’s skin.

• If the pads are not sticking, dry the

patient's chest and shave or clip any
excessive chest hair.

• Reposition the pads.
• Make sure pads are not in the case

or on patient’s clothing.

• Make sure the pads connector is fully

inserted.

If the voice instruction continues after
you do these things, replace the pads
set.