beautypg.com

Fconfiguration, Overview, Device options – Philips 861304 User Manual

Page 49

background image

41

Ph

ilip

s M

e

di

ca

l Sy

st

e

m

s

AP
P

E

ND

IC

ES

F

Configuration

Overview

The Philips HeartStart FRx Defibrillator comes with a factory default configuration
designed to meet the needs of most users. This configuration can only be changed
by using HeartStart Configure version 1.0 or higher, Event Review version 3.2 or
higher, or Event Review Pro 3.1 or higher. This software is for use by trained
personnel. Information about HeartStart data management products is available
online www.medical.philips.com/goto/eventreview. See Appendix A for ordering
information.

Device options

The following table includes the features of FRx operation that are not related to
patient treatment

parameter

settings

default

default description

speaker volume

1, 2, 3, 4,

5, 6, 7, 8

8

The volume of the FRx’s speaker is set to 8,
highest.

auto send periodic

self-test (PST) data

On, Off

On

Enables the periodic self-test data to be
broadcast through the device's infrared
data port.

ECG out data

On, Off

On

Enables the ECG data to be broadcast
through the device's infrared data port.