42 compact 2, Trouble shooting – Protec Protech Compact2 Film Processor SX 2 User Manual
Page 42

42
COMPACT 2
TM
Trouble Shooting
®
7.3 Dryer ventilation is too weak
•
If the ventilation is connected incorrectly, it runs very slowly (heating element
in the air channel starts to glow).
Dryer ventilation connection:
7.4 Dryer temperature cannot be reached
•
Check temperature sensor: Measure voltage at X24 on the power PCB
between pin 3 (green) and pin 2 (brown). The value should read 0.32 V at a
temperature of 32 °C.
•
Only cold air flows from air channel: Heating element in air channel is defec-
tive. Disconnect X12 clip on the power PCB and measure resistance of heat-
ing element (approx. 26
Ω).
•
Enter dryer temperature of 70 °C. Check voltage of X12 on power PCB. If no
voltage can be registered - exchange control unit.
7.5 Dryer temperature display exceeds the set value
•
Check temperature sensor: Measure voltage at X24 on the power PCB
between pin 3 (green) and pin 2 (brown). The value should read 0.32 V at a
temperature of 32 °C.
•
Enter dryer temperature of 35 °C. Check voltage of X12 on power PCB. If no
voltage can be registered - exchange control unit.
7.6 Dryer temperature cannot be reached or is too high
(display shows values approx. 30 °C and 75 °C)
•
Check temperature sensor. Sensor is either not connected or defective.
8
Water
8.1 Rinsing water does not flow, ventilation runs
•
Water pressure in the water system is too low: Minimum pressure 2 bar
(29 psi).
•
Valve activates, no flow passage - filter at inflow is blocked.
•
Check solenoid valve.
8.2 Water tank overflows
•
Water drainage hose (overflow) should have a constant fall. The hose end
should be positioned above the drainage level in the syphon.›‹›‹
•
Check water drainage in the tank and hose for blockage and pollution.
•
When extreme algae growth is noticed, the overflow can be connected
directly at the back of the water tank.
X9
L
black
X9
Z
brown
X10
N
blue
X10
PE
yellow / green