Detailed information about your system

Comparing meter and laboratory results
Test results with the OneTouch

®

UltraMini

Meter are plasma-calibrated. This helps you

and your healthcare professional to compare your meter results with laboratory tests. If
you have been using another type of meter

one that provides whole-blood-calibrated

results

you may notice that your test results with the OneTouch

®

UltraMini

Meter are

approximately 12% higher.

OneTouch

®

UltraMini

Meter test results and laboratory test results both are expressed

in plasma-equivalent units. However, your meter result may differ from your laboratory
result due to normal variation. Meter results can be affected by factors and conditions
that do not affect laboratory results in the same way.

Your OneTouch

®

UltraMini

Meter glucose value is considered accurate when it is

within ±20% of the laboratory measurement. There are some specific situations that
could cause a difference of more than ±20%:

• You have eaten recently. The blood glucose level from a finger stick can be up to

70

mg/dL

higher than blood drawn from a vein (venous sample) used for a lab test.

1