You may have high
glucose and have tested
in an environment near
the low end of the
system’s operating
temperature range
(43–111

°

F).

or,

see top of next page

If you tested in a cool
environment, repeat the test in a
warmer environment with a new
test strip; see pages 22–23. If
this error message appears
again, contact LifeScan Customer
Service at 1 800 227-8862.

Message

What it means

What to do

One of the following may apply:

Under

st

anding

err

or mes

sag

es

Error message indicates
that the blood or control
solution sample was
applied before the meter
was ready.

Repeat the test with a new test
strip. Apply a blood or control
solution sample only after the
flashing blood drop icon

R

appears on the display. If this
message continues to appear,
contact LifeScan Customer
Service at 1 800 227-8862.

DCO_06397301A_OTUM_OB_EN.qxd 05/05/06 9:16 AM Page 40