beautypg.com

Operating instructions, Storing of items – Sanyo BLOOD BANK MBR-304DR User Manual

Page 17

background image

16

OPERATING INSTRUCTIONS

Storing of items

Do not disturb the air circulation in the chamber by the stored items. The disturbed air circulation can

cause the freezing of stored items or degrade of temperature distribution in the chamber.

Air-intake vent

Do not block the air-intake by the stored items near the air-intake vent. Take should be taken not to put

any paper or plastic chips near the vent. They may be vacuumed up and block the vent.

Air-exhaust vent

Do not block the air-exhaust by the stored item near the air-exhaust vent. The stored item may be

frozen if they put nearby the air-exhaust vent.

Load line

The stock area is limited to ensure the air flow in the chamber. The labels

“load line” are attached to the

right and left side of the chamber. Always put the items within the specified area.

Note:

Do not stock any corrosive materials when the container is not sealed completely. Such material causes

corrosion of inner components or evaporator.