9

REFRIGERATOR COMPONENTS

1. Door switch: To detect the door status (open/close).

2. Battery switch: ON-OFF switch for built-in battery for power failure alarm. Always turn on this

switch when the refrigerator is operated. Turn off this switch when the refrigerator is not used for long

period to protect the battery.

3. Power switch: Main switch of the unit.

4. Glow starter: For a fluorescent lamp. It is recommended to replace the glow starter at the time of

replacement of fluorescent lamp. Refer to page 21 for the replacement.

5. Control panel: Panel can be opened when upper right corner of the lower part cover is pushed.

The running status is displayed on this panel and temperature setting is accomplished through this panel.

Refer to page 10 for details.

6. Light switch: This switch is used for turning the fluorescent lamp off and on.

7. Glass window: The window may have condensation in high humidity environment. Wipe off the

condensation with a soft dry cloth.

8. Handle: Always hold the handle when opening/closing the door.

9. Lock: Turn the key counterclockwise through 180

o

to securely lock the door.

10. Fluorescent lamp: 15 W lamp. Normally turn off the lamp. See page 20 for the replacement.

11. Monitor bottle for temperature sensor (2 locations): The bottles has a sensor to display the

chamber temperature. Fill the bottle with 200ml of 10% glycerol or water before commencing the

operation. Refer to page 13 for filling the monitor bottle.

12. Access port: This port allows temperature measurement cables or sensors to enter the chamber

from outside.

13. Shelf (MBR-304D, MBR-304G), Drawer (MBR-304DR, MBR-304GR): Items to be stored in the

chamber must be placed on the shelves or drawers. The maximum storage weight for each shelf or

drawer is 20 kg. Do not put stored items directly onto the interior floor of the chamber.

14. Condensation tray: The condensation on the door is accumulated in this tray. Wipe off the water

occasionally.

15. Caster: Used when moving the unit. When installing the unit the castors can be raised from the

ground by using the leveling feet.

16. Leveling foot: Adjust the height of the leveling foot to install the refrigerator in level. See page 12

for details.

17. Automatic temperature recorder: For details, see the instruction manual of the recorder enclosed

with the unit.

18. Air exhaust vent: Do not block this vent. Care should be taken not to expose the stored items to

direct cooled air.

19. Inner door: Made of acrylic resin. For cleaning, use the diluted neutral dishwashing detergent

20. Air intake vent: Do not block this vent. If this vent is blocked, temperature regulation will become

unstable. Do not insert fingers or any substances into the vent because the circulating fan is running

inside the vent.

21. Door latch: By turning the latch downward, the door is locked. A padlock can be provided

separately.