beautypg.com

Refrigerator components, Control panel and keypad – Sanyo BLOOD BANK MBR-304DR User Manual

Page 11

background image

10

REFRIGERATOR COMPONENTS

Control panel and keypad

1. Alarm buzzer stop key (BUZZER): To silence the audible alarm, press this key.

2. Temperature display key (UPPER/LOWER): By pressing this key, the displayed temperature is

changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature

of upper and lower monitor bottles are displayed in sequence.

3. Alarm test key (ALARM TEST): To check the operation of alarm buzzer, lamp, and remote alarm

terminal.

4. Calibration key (CAL): To calibrate the temperature. This key is used only when the temperature

calibration is necessary.

5. Digit shift key ( ): Pressing this key in the CAL mode will cause the changeable digit to shift.

6. Numerical value shift key ( ): Pressing this key in the CAL mode will cause the numerical value

to change.

7. Door check lamp (DOOR): This lamp is lit when the door is ajar or open.

8. Alarm lamp (ALARM): This lamp blinks during an alarm condition (temperature alarm or power

failure alarm).

9, 10. Temperature display monitor lamp (UPPER, LOWER): Lamp 9 is lit when upper monitor bottle

temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed.

Both 9 and 10 are lit when the average temperature of lower and upper monitor bottle temperature is

displayed.

11. Digital temperature indicator: This indicator shows the current chamber temperature.

1

7 8 9

10

11

2 3 4 5 6