beautypg.com

Owner's record, Warning, Warning on power connection – Sony UP-D23MD User Manual

Page 2: Warning on power connection for medical use

background image

2

Owner's Record

The model and serial numbers are located at the rear.
Record these numbers in the space provided below.
Refer to these numbers whenever you call upon your
Sony dealer regarding this product.

Model No. ____________________
Serial No. ____________________

To prevent fire or shock hazard, do not expose
the unit to rain or moisture.

To avoid electrical shock, do not open the
cabinet. Refer servicing to qualified personnel
only.

THIS APPARATUS MUST BE EARTHED.

Important safeguards/notices for use in
the medical environments

1. All the equipments connected to this unit shall be

certified according to Standard IEC60601-1,
IEC60950, IEC60065 or other IEC/ISO Standards
applicable to the equipments.

2. When this unit is used together with other equipment

in the patient area*, the equipment shall be either
powered by an isolation transformer or connected via
an additional protective earth terminal to system
ground unless it is certified according to Standard
IEC60601-1.
* Patient Area

3. The leakage current could increase when connected

to other equipment.

4. This equipment generates, uses, and can radiate

frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference to other equipment. If this unit causes
interference (which can be determined by

unplugging the power cord from the unit), try these
measures: Relocate the unit with respect to the
susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
Consult your dealer. (According to standard
EN60601-1-2 and CISPR11, Class B, Group 1)

Caution

When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.

Warning on power connection

Use a proper power cord for your local power supply.

Warning on power connection for
medical use

Please use the following power supply cord.
With connectors (plug or female) and cord types other
than those indicated in this table, use the power supply
cord that is approved for use in your area.

*Note: Grounding reliability can only be achieved when the equip-
ment is connected to an equivalent receptacle marked ‘Hospital Only’
or ‘Hospital Grade’.

WARNING

This symbol indicates the equipotential
terminal which brings the various parts of a
system to the same potential.

This symbol is intended to alert the user to
the presence of important operating and
maintenance (servicing) instructions in the
literature accompanying the appliance.

R1.5m

United States

Canada

Plug Type

HOSPITAL GRADE*

HOSPITAL GRADE*

Female end

E62405, E35708

LR53182, LL022442,
LL088408

Cord type

E159216, E35496
Min.Type SJT
Min.18 AWG

LL112007-1, LL20262,
LL32121, LL84494
Min.Type SJT
Min.18AWG

Minimum cord set
rating

10A/125V

10A/125V

Safety approval

UL Listed

CSA