Millenium™ RO System by GE Osmonics

14693e - 31 Mar 03

6 - 14

Drawings

6. TURN ON WATER AND POWER AND THEN CHECK PRESSURE READING. REPEAT AS NECESSARY.

4. USE SCREWDRIVER AND TURN SCREW AS SHOWN. CLOCKWISE TO INCREASE PRESSURE, COUNTER

CLOCKWISE TO DECREASE PRESSURE. EACH 90° ROTATION WILL CHANGE PRESSURE ABOUT 25 PSI.

WASHER

PUMP

ACORN NUT

INCREASE

PRESSURE ADJ. SCREW

DECREASE

PROCEDURE:

TOOLS REQUIRED:

5. REPLACE WASHER (IF IT CAME OFF) AND ACORN NUT.

2. OPEN FRONT COVER AND EXPOSE PUMPHEAD.

3. REMOVE ACORN NUT. NOTE: DO NOT LOSE WASHER.

1. TURN OFF POWER AND WATER TO UNIT.

1. FLAT BLADE SCREWDRIVER.

2. 3/4" WRENCH OR EQUIVALENT.

3. TOWEL.

Millenium RO

(Pump Pressure Adjustment Procedure)