GE OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT User Manual
Page 18

M
Millenium™ RO System by GE Osmonics
14693e – 31 Mar 03
1 - 12
General Information
1.10 Specifications:
1.10.1 Input Water Requirements
Minimum
Maximum
Input
flow
Rate 2.0
GPM
N/A
Temperature
(water)
1.7°C
(35°F)
32°C
(90°F)
NOTE:
Operating at lower temperatures may make it impossible for the Millenium™ to
produce adequate permeate for the dialysis machine.
pH
(operating)
4.0
11.0
TF
Membranes
pH w/chloramines present (operating)
4.0
8.5
TF
Membranes
pH
(cleaning)
2.0
11.0
TF
Membranes
Pressure,
dynamic
30
PSI
100
PSI
1.10.2 Flow
Rates
Product Flow based on the following conditions:
New
Membranes
Operating
Pressure:
190
PSI
Product
Pressure:
1
PSI
Feed Water Temperature:
77
°
F (25
°
C)
GPM LPM GPD LPD
.34 1.31 500 1892.5
.52 1.97 750 2839.1
Membrane flux may vary ±15%
Waste Flow: .84 GPM (3.2 LPM) (
±
10%)
1.10.3 Electrical
Requirements
Motor
1/3 HP
Voltage 115
VAC
/
230
VAC
Hertz
60
/
50
Amps
8
/
5
Phase
single