CAUTION:
‘RENALIN®’ is corrosive to most metals other than stainless steel. Any brass
components that will come into contact with the ‘RENALIN®’ must be replaced with
suitable stainless or plastic materials prior to attempting disinfection. ‘RENALIN®’ type
products contain oxidizing agents and are potentially damaging to thin film (polyamide)
membranes. It is imperative that all the precautions and instructions are followed
correctly. Failure to comply with these instructions, which incorporate instructions from
the membrane manufacturer, will jeopardize the system performance and the warranty.

Renalin®, Minncare™, Peracidin™ and others are trade names for compounds based
on Peracetic Acid and Hydrogen Peroxide compounds, and are common in dialysis
facilities. The products referred to in this procedure typically have 20% hydrogen
peroxide and 4% peroxacetic acid before dilution. Other products are available with
these same chemicals in different ratios and are not acceptable for use in the Z-Series
RO. Actril™ and other compounds with different combinations of these chemicals
must not be used on polyamide (thin-film) membranes.
The name, “Renalin®” will
be used in this manual to refer to acceptable products.

Disinfecting Materials
Description

GE Osmonics Part Number

Disinfection/Cleaning Kit

12214

Renalin®, 1 quart

Not Provided by GE Osmonics

Renalin® Test Kit

Not Provided by GE Osmonics

“DO NOT USE” Warning Labels Not Provided by GE Osmonics
Cleaning/Disinfect Log Sheets

Reference Tech Notes 143, 144, 145

Operation Log Sheet

Reference Tech Note 142

Body Protective Wear

Not Provided by GE Osmonics

1238339b – 16Oct03

4 - 21

Routine Maintenance