beautypg.com

2 calibration procedure 36, Calibration procedure, 2 calibration procedure – Fluke HART SCIENTIFIC 1502A User Manual

Page 39

background image

power. The *C0=, *C1= , and
*C4= commands can be used to set the values of the CAL0,
CAL100, and CAL400 parameters respectively.

8.2

Calibration Procedure

Calibration requires four-wire 100

Ω and 400Ω resistors of 6.25 ppm uncer-

tainty and a 0

Ω resistor (or short). For verification, 25Ω and 200Ω resistors of

10 ppm accuracy are also required. The resistors are connected to the input the
same way probes are. The calibration procedure is as follows:

1.

Connect a 0

Ω resistor to the input and measure its resistance. Note the

average error in the measurement. Adjust the CAL0 parameter by sub-
tracting the measured error. For example, if the input is exactly 0.0000

Ω

and readout shows –0.0011

Ω, the CAL0 parameter should be adjusted

by adding 0.0011 to it.

2.

Connect a 100

Ω resistor to the input and measure its resistance. Note the

average error in the measurement. Adjust the CAL100 parameter by sub-
tracting the measured error. For example, if the input is exactly
100.000

Ω and the readout shows 100.029Ω, the CAL100 parameter

should be adjusted by subtracting 0.029 from it.

3.

Connect a 400

Ω resistor to the input and measure its resistance. Note the

average error in the measurement. Adjust the CAL400 parameter by sub-
tracting the measured error. For example, if the input is exactly
400.000

Ω and the readout shows 399.991Ω, the CAL400 parameter

should be adjusted by adding 0.009 to it.

4.

Verify the accuracy at 0

Ω, 25Ω , 100Ω, 200Ω, and 400Ω. The accuracy

should be within the short-term accuracy limits given in the
specifications.

1502A Thermometer Readout
User’s Guide

36