1 reading temperature 28, Table 6 command list continued 28, Reading temperature – Fluke HART SCIENTIFIC 1502A User Manual
Page 32: Table 6, Command list continued

7.3.1.1
Reading Temperature
The most recent temperature measurement can be read using the following
command:
T
The syntax of the response is as follows:
t:_nnnn.nnn_u
or
t:_nnnn.nnn_u_hh:mm:ss
The _’s represent space characters. The n’s represent the digits of the measure-
ment value. If fewer digits are needed the leading positions are filled with space
characters. The u represents the unit which is either ‘C’, ‘F’, ‘K’, or ‘O’ (for
ohms). The time stamp appears if this option is enabled (see Section 7.3.1.4 be-
low). The time appears in 24-hour format with two digits each for hours, min-
utes, and seconds.
The following SCPI compatible command can also be used to return the most
recent measurement but without the label and unit.
FETC?
FETCH?
1502A Thermometer Readout
User’s Guide
28
Command
Description
Calibration Commands
*PA=
disable password lockout of calibration commands
*LO=[=CA|AL]
read [or set] menu lockout
*C0[=
read [or set] the 0
Ω calibration parameter
*C1[=
read [or set] the 100
Ω calibration parameter
*C4[=
read [or set] the 400
Ω calibration parameter
*SN[=
read [or set] the instrument serial number
Miscellaneous Commands
*VER
read model number and firmware version number
*IDN?
read manufacturer, model number, serial number, and firmware version number (SCPI
compatible)
H
read a list of commands
Command List Continued