LF=OF
7.3.5
Calibration Commands
The following commands are used in calibrating the instrument.
7.3.5.1
Entering the Password
In order to set the calibration parameters the password must be issued first. The
following command enables access to the calibration parameters:
*PA=2051
Calibration parameters can be locked out again by sending *PA=0 or by cycling
the power.
7.3.5.2
Setting the Menu Lockout
The following commands can be used to select the menu lockout options:
*LO=CA
*LO=AL
7.3.5.3
Setting the Calibration Coefficients
The instrument calibration coefficients are used to maintain the resistance mea-
surement accuracy of the 1502A. These coefficients must not be changed ex-
cept by a qualified technician during the calibration of the 1502A. The
following commands can be used to set the instrument calibration coefficients:
*C0=
*C1=
*C4=
7.3.5.4
Setting the Serial Number
The following command is used to set the serial number of the 1502A:
*SN=
7.3.6
Other Commands
Remaining commands are described below.
7.3.6.1
Instrument Identification
The following command returns the model number and firmware version
number:
*VER
The syntax of the response is as follows:
1502A Thermometer Readout
User’s Guide
32