beautypg.com

Front panel lock out conditions, Front panel review only function, Front panel monitor only function – Fluke 2625A User Manual

Page 68

background image

2620A, 2625A

Users Manual

3-26

Table 3-17. Autoprint/Memory Storage Selection

Destination

(“dESt”) (Note

1)

Mode (“MOdE”)

(Note 2)

Press these
buttons:P

MODE

KU

)

G

D

E

G

D

E

To select from
these choices:

Print

StorE

both

ALL

ALAr

trAnS

Note 1.

“Print”

Sends data to be printed through the RS-232 interface.

“StorE”

Sends data to be stored in memory. (Hydra Data Logger only.)

“both”

Sends data to be stored and printed (Hydra Data Logger only.)

Note 2.

“ALL”

Measurements for all defined channels are printed/stored when a scan occurs.

“ALAr”

Measurements for all defined channels are printed/stored when a scan occurs
and at least one channel is in alarm.

“trAnS”

Measurements for all defined channels are printed/stored when a scan occurs
and at least one channel has transitioned into or out of alarm since the last
scan.

Table 3-18. Clearing Memory Storage

To Initiate

To Confirm

To Activate

Hold

K

, and press

U

“no” to exit without clearing

G

D

E

“CLEAr” appears in right display

“YES”

C

to exit without clearing

Front Panel Lock out Conditions

Various methods are available to prevent accidental use of the front panel buttons. These
actions can be initiated from either the front panel or the computer interface.

Front Panel Review Only Function

Access the Review Array (press

N, REVIEW annunciator comes on)

Press

F and B simultaneously to activate the Review Only function (REM

annunciator comes dim).

J, H, G, and D can now be used to view the various elements of the Review Array.
All other front panel buttons are locked out, yielding a long beep when pressed. Press
both

F and B again to deactivate the Review Only function and return the

instrument to normal front panel button operation (regular Review Array display.)

Front Panel Monitor Only Function

First place the instrument in the Monitor Function (press

M, MON annunciator comes

on). Then press

F and B simultaneously to activate the Monitor Only Function

(REM annunciator comes on).

This manual is related to the following products: