Fluke 2625A User Manual

Page 134

background image

2620A, 2625A

Users Manual

6-6

Table 6-2. Recommended Test Equipment

Instrument Type

Minimum Specification

Recommended

Model

Multifunction Calibrator

DC Voltage:

Fluke 5700A

Range: 90 mV to 300V dc
Accuracy: .005%

AC Voltage:

Frequency

Voltage

Accuracy

1 kHz

29 mV to 300V

0.05%

100 kHz

15 mV to 300V

1.25%

Frequency:

10 kHz

1V rms

.0125%

Decade Resistance
Source

Ohms

Accuracy

General Resistance
Inc. Model RDS 66A

290

0.0125%

2.9 k

0.0125%

29 k

0.0125%

290 k

0.0125%

2.9 M

0.0125%

Note

The 5700A Calibrator can be used for 0.05%
accuracy (rated) on the 3.0 k

, 30 k

, and 3.0

M

ranges. The 5700A can be used for 0.06%

accuracy on the 300

range.

Mercury Thermometer

0.02 degrees Celsius resolution

Princo ASTM-56C

Thermocouple Probe

Type K

Fluke P-20K

Room Temperature
Oil/Water Bath

Thermos bottle and cap

Multimeter

Measures +5V dc

Fluke 77

Signal Generator

Sinewave, 0.5 to 1V rms 10 Hz to 5 kHz

Fluke 6011A

Alternate Equipment List

(Minimum specifications are the same as in the Standard Equipment List)

Instrument Type

Recommended Model

DC Voltage Calibrator

Fluke 5440B

DMM Calibrator

Fluke 5100B (for AC Volts only)

Function/Signal Generator

Philips PM5193 or Fluke 6011A