Frequency Inputs

Frequency Range

15 Hz to greater than 1 Mhz

Range

Resolution

Accuracy + (%

±

Hz)

Slow

Fast

Slow

Fast

15 Hz - 900 Hz

0.01 Hz

0.1 Hz

0.05% + 0.02 Hz

0.05% + 0.2 Hz

9 kHz

0.1 Hz

1 Hz

0.05% + 0.1 Hz

0.05% + 1 Hz

90 kHz

1 Hz

10 Hz

0.05% + 1 Hz

0.05% + 10 Hz

900 kHz

10 Hz

100 Hz

0.05% + 10 Hz

0.05% + 100 Hz

1 MHz

100 Hz

1 kHz

0.05% + 100 Hz

0.05% + 1 kHz

Sensitivity

Frequency

Level (Sine Wave)

15 Hz - 100 kHz

100 mV rms

100 kHz - 300 kHz

150 mV rms

300 kHz - 1 MHz

2 V rms

Above 1 MHz

Not specified

Maximum AC Input

300V rms or 424V peak on channels 0, 1, and 11
150V rms or 212V peak on channels 2 to 10 and 12 to 20
Voltage ratings between channels must not be exceeded

2 x 10

6

Volt-Hertz product on any range, normal mode input

1 x 10

6

Volt-Hertz product on any range, common mode input

Cross-Talk Rejection

Refer to Appendix D.