Fluke 2625A User Manual
Page 102

2620A, 2625A
Users Manual
4-32
Table 4-8. Command and Query Reference (cont)
MEASUREMENT
UNITS STRING
Scaled “MX+B”
Volts DC
“VDC”
Volts AC
“VAC”
Resistance
“OHMS”
Frequency
“Hz”
Temperature
“C”
Temperature
“F”
FORMAT?
Query Response Format
Returns the format presently in use:
The default. IEEE-488.2-compatible. No measurement units are allowed.
Responses are RS-232-compatible, allowing a means of recording the
units with the measurement value.
FUNC
Channel Function Definition
Define the measurement function and range for the indicated channel.
Attempting to use the FUNC command when measurements are active
results in an Execution Error.
Note
Successful execution of the FUNC command clears any alarm limits and
scaling values for this channel. Therefore, you must define a channel’s
function before setting alarm limits and/or scaling values for that channel.
Also, the FUNC command clears all values in the Review array. FUNC
FUNC
FUNC
FUNC
FUNC
If the channel number is invalid, an Execution Error is generated.
If the function is OFF, any additional supplied parameters generate a
Command Error.
Ranges for VAC, VDC, OHMS, and FREQ are listed in the following
table. Enter AUTO to select autoranging. Use of any other values causes
an Execution Error.
RANGE
VOLTAGE
OHMS
FREQUENCY
1
300 mV
300
Ω
900 Hz
2
3 V
3k
Ω
9 kHz
3
30 V
30 k
Ω
90 kHz
4
150/300 V
300 k
Ω
900 kHz
5
Error
3 M
Ω
1 Mhz
6
Error
10 M
Ω
Error